Measuring Instruments

CRT Color Analyzer
Higher levels of accuracy, speed, sensitivity, and convenience
| CA-100Plus (Measuring Probe) | CA-100Plus (High Luminance Measuring Probe) | ||
|---|---|---|---|
| Detector | Silicon photocell | Silicon photocell | |
| Measurement diameter | Ø 21 mm | Ø 21 mm | |
| Display range | Luminance | 0.01 to 1000 cd/m2 | 0.01 to 2000cd/m2 |
| Chromaticity | Displayed in 4- or 3-digit values (selectable) | Displayed in 4- or 3-digit values (selectable) | |
| Luminance | Measurement range | 0.05 to 1000 cd/m2 | 0.05 to 2000 cd/m2 |
| Accuracy (for white)*1 | ±2%, ±1 digit of reading (Temperature: 23°C±2°C, Relative humidity: (40±10)%) | ±2%, ±1 digit of reading (Temperature: 23°C±2°C, Relative humidity: (40±10)%) | |
| Repeatability*1 | 0.05 to 1000 cd/m2 0.2%+1 digit (2σ) |
0.05 to 2000 cd/m2 0.2%+1 digit (2σ) |
|
| Chromaticity | Measurement range | 0.05 to 1000 cd/m2 | 0.05 to 2000 cd/m2 |
| Accuracy*1 | 0.05 to 0.19 cd/m2 ±0.006 for white; 0.20 to 0.49 cd/m2 ±0.004 for white; 0.50 to 1000 cd/m2 ±0.003 for white; 40.00 cd/m2 ±0.002 for white, ±0.004 for monochrome*2 | 0.05 to 0.09 cd/m2 ±0.008 for white; 0.1 to 0.39 cd/m2 ±0.006 for white; 0.40 to 0.99 cd/m2 ±0.004 for white; 1.00 to 2000 cd/m2 ±0.003 for white; 40.00 cd/m2 ±0.002 for white, ±0.004 for monochrome*2 | |
| Repeatability*1 | 0.05 to 0.19 cd/m2 0.006 (2σ); 0.20 to 0.49 cd/m2 0.002 (2σ); 0.50 to 1000 cd/m2 0.001 (2σ) | 0.05 to 0.09 cd/m2 0.009 (2σ); 0.1 to 0.39 cd/m2 0.006 (2σ); 0.40 to 0.99 cd/m2 0.002 (2σ); 1.00 to 2000 cd/m2 0.001(2σ) | |
| Measurement speed*3 | xyLv | 0.05 to 0.99 cd/m2 5 (4.5) meas./sec.*4 ; 1.00 to 1000 cd/m2 20 (17) meas./sec.*4 | 0.05 to 1.99 cd/m2 5 (4.5) meas./sec.*4 ; 2 to 2000 cd/m2 20 (17) meas./sec.*4 |
| Display | Digital | xyLv, TΔuvLv, RGB, XYZ, u'v'Lv | xyLv, TΔuvLv, RGB, XYZ, u'v'Lv |
| Analog | Δ x, Δ y, Δ Lv, R/G, B/G, Δ G, Δ R, B/R, G/R | ||
| SYNC mode | NTSC, PAL, EXT, UNIV, INT | NTSC, PAL, EXT, UNIV, INT | |
| Object under measurement | Vertical synchronous frequency: 40 to 200Hz | Vertical synchronous frequency: 40 to 200Hz | |
| Memory channels | 100 | 100 | |
| Analyzer function | Standard function | Standard function | |
| Interface | USB (Version 1.1) or RS-232C (38,400bps or below) | USB (Version 1.1) or RS-232C (38,400bps or below) | |
| Multi-point measurement | Max. 5 points (Use 4-Probe Expansion Board CA-B04.) | Max. 5 points (Use 4-Probe Expansion Board CA-B04.) | |
| Operating temperature/humidity range | Temperature: 10 to 28°C; Relative humidity 70% or less with no condensation Luminance change: ±2% ±1 digit of reading for white Chromaticity change: ±0.002 for white, 0.006 for monochrome (from reading of Konica Minolta's standard CRT*1, 40.00cd/m2, at 23°C, 40% | Temperature: 10 to 28°C; Relative humidity 70% or less with no condensation Luminance change: ±2% ±1 digit of reading for white Chromaticity change: ±0.002 for white, 0.006 for monochrome (from reading of Konica Minolta's standard CRT*1, 40.00cd/m2, at 23°C, 40% | |
| Storage temperature/humidity range | 0 to 28°C: Relative humidity 70% or less with no condensation 28 to 40°C: Relative humidity 40% or less with no condensation | 0 to 28°C: Relative humidity 70% or less with no condensation 28 to 40°C: Relative humidity 40% or less with no condensation | |
| Input voltage range | 100 to 240V~, 50 to 60Hz, 50VA | 100 to 240V~, 50 to 60Hz, 50VA | |
| Dimensions (W × H × D) | Main unit, Probe | 340 × 127 × 216 mm, Ø 45 × 142 mm | 340 × 127 × 216 mm, Ø 45 × 142 mm |
| Weight | Main unit, Probe | 3.58 kg, 285 g | 3.58 kg, 285 g |