Illuminace Meter T-10A

A professional illuminance meter that provides accurate and traceable lux or fcd over a large measuring range. Can be used with a detachable receptor head for remote measurement or in multi-point mode by connecting up to 30 receptors to cover a large area.

Specifications

Principal Specifications T-10A

Model

Illuminace Meter T-10A

Type Multi-function digital illuminance meter with detachable receptor head
Illuminance meter class Conforms to requirements for Class AA of JIS C 1609-1: 2006
"Illuminance meters Part 1: General measuring instruments"
Conforms to DIN 5032 Part 7 Class B
Receptor Silicon photocell
Relative Spectral Response* Within 6% (f1´) of the CIE spectral luminous efficiency V (λ)
Cosine Correction Characteristics Within ±1% at 10°; Within ±2% at 30°; Within ±6% at 50°; Within ±7% at 60°; Within ±25% at 80°
Illuminance units Lux (lx) or foot candles (fcd) (switchable)
Measuring range Auto range (manual 5 range at the time of analog output)
Light source Illuminance (lx). illuminance difference (lx). illuminance ratio (%). integrated illuminance (lx·h). integration time (h). average illuminance (lx).
Measuring range Illuminance 0.01 to 299,900 lx 0.001 to 29,990 fcd
Integrated illuminance 0.01 to 999,900 × 103 lx·h 0.001 to 99,990 × 103 fcd·h / 0.001 to 9999 h
User calibration function CCF (Colour Correction Factor) setting function
Accuracy ±2% ±1digit of displayed value (based on Konica Minolta standard)
Temperature / humidity drift Within ±3% ±1digit (of value displayed at 20°C/68°F ) within operating temperature/humidity range
Digital output USB
Analog output 1mV/digit,3V at maximum reading; Output impedance: 10KΩ; 90% response time: FAST setting: 1ms, SLOW setting: 1s
Operating temperature/humidity range -10 to 40°C, relative humidity 85% or less (at 35°C) with no condensation
Storage temperature/humidity range -20 to 55°C, relative humidity 85% or less (at 35°C) with no condensation
Power source 2 AA-size batteries / AC adapter (optional)
Battery life 72 hours or longer (when alkaline batteries are used) in continuous measurement
Standard accessories White calibration plate CM-A 133, Zero calibration box CM-A119, AC adapter AC-A12, RS-232C cable IF-A12, Accessory Case CM-A117, Dust Cover CM-A118, Unit Driver CM-A108
Optional accessories Receptor head ; Adapter for Multi-point ; AC Adapter ; Data processing software

Specifications are subject to change without prior notice.