Imaging colorimeters for 2D light measurement
Solution for complex measurement requirements in aviation, automotive, general lighting or the display industry.
Principal Specifications
Principal Specifications of the LumiCam 2400B Series
LumiCam 2400B | Mono | Color | Advanced | |
Quantities | ||||
Photometric quantities | Luminance (cd/m2), luminous intensity (cd), contrast | |||
Colorimetric quantities | - | Color coordinates (x,y), color coordinates (u',v'), tristimulus values (X, Y, Z), dominant wavelength (nm), color saturation, correlated color temperature CCT (K) | ||
Camera data | ||||
Sensor | Sony IMX264LLR CMOS Sensor | |||
Sensor size | 2/3’’, 11.1 mm diagonal | |||
Effective number of pixels (h x v) | 2428 x 2028 (5 MP) | |||
Pixel size | 3.45 µm x 3.45 µm | |||
AD converter | 12 bit | |||
Exposure time | 40 µs to 30 s | |||
Luminance measurement | ||||
Measurement range | 0.3 mcd/m² … 2.5 mcd/m² | |||
Extended measurement range | 2.5 x 1010 cd/m² | |||
Measurement time incl. data saving time | 1.9 s | |||
Measurement time incl. data saving time | 1.8 s | |||
Accuracy for std. illuminant A | ±3 % | |||
Accuracy for LED color light | - | - | ±2 % | |
Repeatability | ±0.03 % | |||
Uniformity | ±0.5 % | |||
Filter match | f1' < 3 % | |||
Color measurement | ||||
Measurement time incl. data saving time | - | 10 s | 16 s | |
Measurement time incl. data saving time | - | 8 s | 12 s | |
Accuracy (x, y) for std. illuminant A | - | ±0.003 | ±0.003 | |
Accuracy (x, y) for color light | - | ±0.010 | ±0.010 | |
Accuracy (x, y) for LED color light | - | - | ±0.005 | |
Repeatability (x, y) | - | ±0.0001 | ±0.0003 | |
Uniformity (x, y) | - | ±0.001 | ||
Filter match | f1' (Y) < 3 % | f1' (Xb) < 6 %, f1' (Xr) < 6 %, f1' (Y) < 3 %, f1' (Z) < 4 % | ||
General | ||||
Interface | Gigabit Ethernet | |||
Operating system | Windows 7 (32/64 bit), Windows 10 (64 bit) | |||
Dimensions incl. objective lens | 210x105x100 mm | 235x119x133 mm | ||
Weight | 1.5 kg | 3 kg | ||
Power supply | 24 V external | |||
Operating conditions | 10 to 40 °C, max. 70 % RH (non-condensing) |
Instrument Systems continuously endeavours to develop and improve all products. Any technical changes, errors or misprints do not form ground for compensation. The company’s terms of delivery and payment apply in all other respects. If not otherwise stated explicitly, all specifications refer to the Nikon 50 mm lens with a 16 mm aperture. The detailed measurement conditions under which the specified values have been determined can be looked up in the data sheets and brochures or requested from us.