Konica Minolta

Europe

Specifications

Luminance & Colour Meter CS-200

The CS-200 Chroma Meter measures luminance and chromaticity of light emitting products as diverse as large plasma displays, compact LCDs and LEDs, outdoor screens, high pressure lamps, and instrument panels.

Specifications

Principal Specifications of the CS-200

Model

Luminance & Color Meters CS-200

Measurement range

0.01 - 200,000 cd/m2 (Measuring angle 1°)

0.01 - 5,000,000 cd/m2 (Measuring angle 0.2°)

0.01 - 20,000,000 cd/m2 (Measuring angle 0.1°)

Accuracy
(Measuring angle 1°) #1
(Temperature: 23°C±2°C, Relative humidity: 65% max.)

150 cd/m2 ( for Illuminant A ) : Lv ±2 % ±1digit ; xy ±0.002

0.01-0.5 cd/m2 ( for Illuminant A ) : Lv ±0.02 cd/m2 ±1digit ;

0.5-1 cd/m2 ( for Illuminant A ) : Lv ±0.02 cd/m2 ±1digit ; xy ±0.007

1-10 cd/m2 ( for Illuminant A ) : Lv ±2 % ±1digit ; xy ±0.004

10-200,000 cd/m2 ( for Illuminant A) : Lv ±2 % ±1digit ; xy ±0.003

5000 cd/m2 ( for Illuminant A ) : color filter (R, G, B) ; xy ±0.006

Repeatability
(Measuring angle 1°) #2

0.01-1 cd/m2 ( for Illuminant A) : Lv ±0.01 cd/m2 ±1digit ;(2σ/AUTO)

1-2 cd/m2 ( for Illuminant A) : Lv ±0.5 % ±1digit ; xy 0.002 (2σ/AUTO)

2-4 cd/m2 ( for Illuminant A) : Lv ±0.5 % ±1digit ; xy 0.001 (2σ/AUTO)

4-8 cd/m2 ( for Illuminant A) : Lv ±0.5 % ±1digit ; xy 0.0005 (2σ/AUTO)

8-200,000 cd/m2 ( for Illuminant A) : Lv ±0.1% ±1digit ; xy 0.0004 (2σ/AUTO)

Measurement time

Automatic setting between 1s and 60s (AUTO); Automatic setting to 1s or 3s

(LTD. AUTO) 0.5 sec / meas.

(Super-FAST) 1 sec / meas.

(FAST)3 sec / meas.

(SLOW) 12 sec / meas.

(Super-SLOW) (approx. 12 sec / meas)

Measurement method

Spectral method, Grating + linear photo diode array

Measuring angle

1°, 0.2°, 0.1°(switchable)

Minimum measuring area

0.5 mm

0.1 mm (close up lens)

Minimum measuring distance

296 mm (Distance from front edge of metal lens barrel)

Observer

2/10 degrees

Colour space

Lv x y, Lv u' v', Lv TΔuv, XYZ, dominant wavelength

Measurement synchronization setting range

Vertical synchronization frequency : 40 to 200Hz

Interface

USB 1.1

Power source

AC adapter or 4 AA-Size Batteries

Battery life

Approx. 3 hours (continuous measurement / Fast mode / AA-size alkaline cells)

Size

95 mm (W) × 127 mm (H) × 330 mm (L)

Weight

1.8 kg (without battery)

Operating temperature / humidity range

0°C to 40°C, relative humidity 85% or less (at 35°C) with no condensation

Storage temperature / humidity range

0°Cto 45°C,relative humidity 85% or less (at 35°C)with no condensation

 #1
23°C±2°C Lv = 0.01-10 cd/m2, SLOW, average of 30 measurements
Lv = 10 cd/m2 and higher, SLOW, average of 10 measurements

#2
At 0.2°measuring angle, the amount of received light is approx. 1/25 of that for 1°.
Therefore, the repeatability becomes the same as that for 1°with 25 times lower luminance.
At 0.1°measuring angle, the amount of received light is approx. 1/100 of that for 1°.
Therefore, the repeatability becomes the same as that for 1°with 100 times lower luminance. 

 

*Specifications are subject to change without prior notice.