Spectroradiometer CS-2000

A high end spectroradiometer for light sources and displays that measures spectral data, luminance and chromacity. Fast measurement times with high repeatability ideal for the low luminance blacks and contrast measurements on modern displays.

Specifications

Principal Specifications CS-2000

Model Spectroradiometer CS-2000
Measurement angel 0.1°, 0.2°, 1° selectable
AMeasurement wavelength range 380 to 780nm
Wavelength resolution 0.9nm / pixel
Display wavelength pitch 1.0 nm
Median wavelength precision ±0.3nm (at 435.8nm, 546.1nm, and 643.8nm)
Half bandwidth 5nm or less
Luminance measurement range (for Standard illuminant A) 0.1° measurement angle: 0.3 to 500,000 cd/m2

0.2° measurement angle: 0.075 to 125,000 cd/m2

1° measurement angle: 0.003 to 5,000 cd/m2
Accurancy Luminance±2%

Chromaticity

(1° measurement

angle)

x,y: ±0.003 (0.003 to 0.005 cd/m2)

x,y: ±0.002 (0.005 to 0.05 cd/m2)

x,y: ±0.0015 (0.05 cd/m2 or more)
Repeatability Luminance (1°C measurement angle)

0.4% (0.003 to 0.05 cd/m2)

0.3% (0.005 to 0.1 cd/m2)

0.15% (0.1 to 5,000 cd/m2)
Chromaticity (1° measurement angle)

x: 0.001 y: 0.0015 (0.003 to 0.1 cd/m2)

x: 0.0006 y: 0.0006 (0.1 to 0.2 cd/m2)

x: 0.0004 y: 0.0004 (0.2 to 5,000 cd/m2)

Polarization error

0.1°, 0.2° measurement angle: 3% or less
1° measurement angle: 2% or less

Measurement time Approx. 1 sec. (Manual mode) to 243 sec.
Colour space mode Lv x y, Lv u’ v’, Lv TΔuv, XYZ ,spectral waveform, dominant wavelength, excitation purity scotopic luminosity (with CS-S10w Professional)
Interface USB 1.1
Operating temperature/humidity range 5 to 35°C; Relative humidity 80% or less with no condensation
Power supply AC adapter (100 to 240V~, 50/60 Hz)
Dimensions (W × H × D) 158 × 200 × 300mm (Main unit) Ø70 × 95mm (Lens)
Weight 6.2kg (including lens)
  • Measurement conditions for accuracy/repeatability: Measurement subject: Standard Illuminant A; Ambient temperature: 23°C±2°C;
  • Relative humidity: 65% or less
  • Specifications and appearance subject to change without notice