Specifikationer
Benchtop Spectrophotometer CM-36d | ||
---|---|---|
Illumination/viewing system | Reflectance: di:8°, de:8° (diffused illumination, 8° viewing) Conforms to CIE No.15 (2004), ISO7724/1, ASTM E1164, DIN 5033 Teil7 and JIS Z 8722 Condition c standards. | |
Integrating sphere size | Ø152 mm | |
Detector | Dual 40-element silicon photodiode array | |
Spectral separation device | Planar diffraction grating | |
Wavelength range | 360-740 nm | |
Wavelength pitch | 10 nm | |
Half bandwidth | Approx. 10 nm | |
Measurement range | 0-200% | |
Resolution: 0.01% | ||
Light source | Pulsed Xenon lamp 1x | |
Illumination / Measurement area [mm] | LAV: Ø30.0 / Ø25.4 MAV: Ø11.0 / Ø8.0 SAV: Ø7.0 / Ø4.0 | |
Measurement time | Approx. 3.5sec (SCI + SCE measurement) | |
Min. measurement interval | Approx. 4sec (SCI + SCE measurement) | |
Repeatability | Colorimetric values: Std. dev. within ΔE*ab 0.03 Spectral reflectance: Std. dev. within 0.1% (When a white calibration plate measured 30 times at 10sec intervals after white calibration) | |
IIA | Within ΔE*ab 0.15 (LAV/SCI) Average for 12 BCRA Series II color tiles compared to values measured with master body. | |
UV control | No adjustment function (UV100%) | |
Sample preview | RGB Camera | |
Internal Performance Check* | WAA (Wavelength Analysis & Adjustment) Technology | |
Interfacing | USB 2.0 | |
Operation temperature / humidity range | 13-33°C / rH 80% or less (at 33°C) with no condensation | |
Storage temperature / humidity range | 0-40°C / rH 80% or less (at 35°C) with no condensation | |
Size [mm] | 248 x 250 x 498 mm | |
Weight [kg] | 8.3 kg |
*WAA license purchase required.
Specifikationer kan ändras utan föregående meddelande.
Företagsnamn och produktnamn är varumärken eller registrerade varumärken som tillhör respektive företag.