Konica Minolta

Europe

Specifications

ProMetric I-Series: 2D Luminance and Color Imaging

Uniquely offering High Resolution, High Speed and High Accuracy, all at once – enabling advanced applications in Automotive, Aerospace, Consumer Electronics, Lighting, Research & Development.

Specifications

Parameter

ProMetric I2

ProMetric I8

ProMetric I16

ProMetric I29

Primary Application

Uniformity Testing, NFMS, ProMetric in R&D Setting

Production Line Testing

CCD Pixels

1600 x 1200

3296 x 2472

4896 x 3264

6576 x 4384

CCD Megapixels

1.9

8.1

16.0

28.8

CCD Type

Interline, Cooled to +5° C.”

System Dynamic Range
(single exposure, per pixel)

 

61.4 dB (1 x 1 binning)

73.4 dB (2 x 2 binning)

High Dynamic Range
(multi-exposure)

>1,000,000:1

Luminance (Minimum)

0.00001 cd/m2 Limit of Detection

0.0001 cd/m2@ SNR = 60

0.0005 cd/m2@ SNR = 100

Luminance (Maximum)

1010 cd/m2 with optional ND filters

System Accuracy*

Illuminance ± 3%; Luminance (Y) ± 3%; Color Coordinates (x,y) ± 0.003

Short-term Repeatability**

Illuminance ± 0.02%; Luminance (Y) ± 0.02%; Color Coordinates (x,y) ± 0.00005

Lens Type / Focal Distances Available

Electronically controlled focus and aperture; 24, 35, 50, 100, 200 mm

Electronically controlled focus and aperture; 35, 50, 100, 200 mm

Electronically controlled focus and aperture; 50, 100, 200 mm

Field of View
(Full Angle, H x V degrees)

24 mm 20° x 15°

35 mm 14° x 10°

50 mm 10° x 8°

100 mm 5° x 4°

200 mm 3° x 2°

24 mm 38° x 30°

35 mm 29° x 22°

50 mm 21° x 16°

100 mm 10° x 8°

200 mm 5° x 4°

35 mm 41° x 28°

50 mm 30° x 20°

100 mm 15° x 10°

200 mm 8°x 5°

50 mm 40° x 28°

100 mm 20° x 14°

200 mm 11° x 7°

Minimum Measurement Time1

0.3 sec - photopic 1.1 sec - color

0.4 sec - photopic 1.2 sec - color

0.6 sec - photopic 1.5 sec - color

0.9 sec - photopic

2.4 sec – color

Spatial Measurement Capabilities

Luminance, Radiance, Illuminance, Irradiance, Luminous Intensity, Radiant Intensity, CIE Chromaticity Coordinates, L*a*b* Color Scale, Correlated Color Temperature (CCT), Dominant Wavelength

Units

foot-lambert, cd/m2, nit, W/sr/m2, foot-candles, lux, lux-s, W/m2, W-s/m2, candela, W/sr.

CIE (x, y) and (u’, v’), Kelvin (CCT)

Communication Interface

Ethernet 100/1000, USB 2.0

Power

100-240 V, 50-60 Hz, 140 Watts

LCD Touch Panel

Resolution: 800 x 600; Diagonal: 125 mm

Dimensions
(H x W x D)

238 mm x 181 mm x 230 mm

Weight

4.9 kg

Operating Temperature

0 - 30° C

Operating Humidity

20 - 70% non-condensing

Specifications subject to change without notice.

* Based on illuminant A or user calibration for specific spectra. Based on a virtual detector size of 1% of the FOV.

** Based on a virtual detector size of 1% of FOV.